| Model Number |
SUTURELASSO 25° RIGHT |
| Medical Device Problem Codes |
Defective Device (2588); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a ramp suture surgery the quicklasso are obstructed and any suture cannot go out of our device.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint is confirmed.Visual inspection did not find any issues with the device returned.Functional testing with suture 2-0 did not go through the shaft to the tip of the 25°, curve, left quickpass¿ lasso, low profile.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force used during suture passing/tightening /tensioning.
|
| |
|
Search Alerts/Recalls
|