| Catalog Number |
362420 |
| Medical Device Problem Code |
Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported by the user facility: brief inquiry description: tubing causing air in line alarms.Detailed inquiry description: tubing causing air in line and upstream occlusion alarms require the nurses to send the drugs back to pharmacy to switch them out because they are chemo regimens.No injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).No sample number was provided for evaluation.Although the defect could not be confirmed, corrective actions were performed to revise specifications which limit unit-to-unit variation in tubing outer diameters and increased the length of the tubing that contacts the air sensor.The reduced variation in tubing outer diameter and length revision improves the coupling between the administration set and the air-in-line sensor reducing the potential for false air-in-line alarms.B.Braun medical inc.Has initiated a voluntary medical device correction 2523676-9/26/23-004-c for multiple batches of infusomat® pump sets manufactured between 01.Jan.2022 and 17.Aug.2023.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|
Search Alerts/Recalls
|