EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PHYSIO II RING; RING, ANNULOPLASTY Back to Search Results
Model Number 5200M
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Mitral Valve Insufficiency/ Regurgitation (4451)
Date of Event 09/12/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was learned from the implant patient registry that a 34mm mitral ring was explanted and replaced with a 30mm mitral ring after an implant duration of 1 year, 9 months due to regurgitation.The perceived root cause of the event is due to structural deficiency/damage of the annuloplasty ring.The patient was stable in the icu post procedure.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital and patient family members) and is not received in the form of a conventional "customer complaint." the information reported may or may not be related to the edwards device.
 
Additional Manufacturer Narrative
There is evidence of device-related harm or structural deficiency of the ring that could cause or contribute to a serious injury.The device was not returned for evaluation, as device request is still ongoing.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
 
Additional Manufacturer Narrative
The most likely cause is patient factors, including progression of the patients valvular disease.The device history record (dhr) was reviewed and showed that this device met all manufacturing and sterilization specifications for product release prior to distribution.No issues were identified that would have impacted this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PHYSIO II RING
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
one edwards way
irvine, CA 92614
9492506615
MDR Report Key18088082
Report Number2015691-2023-17309
Device Sequence Number17839687
Product Code KRH
UDI-Device Identifier00690103180220
UDI-Public(01)00690103180220(17)261012
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K083470
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number5200M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2023
Supplement Date Received by Manufacturer12/04/2023
Initial Report FDA Received Date11/07/2023
Supplement Report FDA Received Date12/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization; Life Threatening;
Patient Age59 YR
Patient SexMale
-
-