| Model Number |
ORT300 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The imris customer service engineer investigated the issue on-site and observed a small hydraulic fluid leak from a pressurized input hose associated with the back lift cylinder.The cse measured the hydraulic fluid level in the operating room table and confirmed it was within specification.The table was functionally tested and no functional issues were present or previously reported by the customer.Replacement of the hydraulic hose is in process; a follow-up report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
A biomedical technician at the customer site reported observing hydraulic fluid at the base of the ort300 operating room table.There was no patient or procedural involvement at the time of the observation.
|
| |
|
Additional Manufacturer Narrative
|
|
The imris service engineer replaced flow control valves and additional hydraulic hoses and fittings in the hydraulic system adjacent to the flow control valves to address the reported fluid leak.Hydraulic fluid levels and system pressure were measured during the service visits and confirmed to be within specification.No table functional impact was reported by the customer or observed by the service engineer throughout multiple site visits, on-site testing, and followup preventive maintenance visits.This report also encompasses udi related data quality updates to match brand name and udi with gudid data.
|
| |
|