| Model Number |
3300 |
| Medical Device Problem Code |
Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a pacing system analyzer (psa) test while performing an implant procedure, the physician experienced difficulties operating this programmer.The physician noted that the programmer display screen was intermittently unresponsive turning off and on during a threshold test.Technical services (ts) recommended to the physician to use a different programmer to rule out possible electromagnetic interference (emi).The programmer remains in service.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for product analysis because the user was able to resolve the issue and the programmer remains in service at the healthcare facility.Although the device was not returned for analysis, boston scientific performed an investigation with all available information.An unresponsive touchscreen is a known and anticipated potential hazard that can occur during use of the device which is accounted for in device risk documentation and labeling.Boston scientific has received similar reported events where the latitude programmer screen became unresponsive to touch and a complaint review confirmed that the occurrence of harm associated with these events is low.Investigation of this behavior has not found any commonalities (i.E., device serial numbers, date of the event, date of manufacture, etc.) in the reports received.The investigation determined that the cause of touchscreens becoming unresponsive during use was due to the water detection feature within the programmer's liquid crystal display (lcd) touchscreen assembly.Various inputs can induce false positive detections of this water detection feature, which then results in an unresponsive touchscreen.
|
| |
|
Event or Problem Description
|
|
It was reported that during a pacing system analyzer (psa) test while performing an implant procedure, the physician experienced difficulties operating this programmer.The physician noted that the programmer display screen was intermittently unresponsive turning off and on during a threshold test.Technical services (ts) recommended to the physician to use a different programmer to rule out possible electromagnetic interference (emi).The programmer remains in service.No patient involvement.
|
| |
|
Search Alerts/Recalls
|