• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. NIM-NEURO® 3.0 MAINFRAME; STIMULATOR, NERVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. NIM-NEURO® 3.0 MAINFRAME; STIMULATOR, NERVE Back to Search Results
Model Number 8253402
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/11/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that pre-op, the current was not delivered during stimulation.There is no patient involved in this event.
 
Additional Manufacturer Narrative
H3:continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 8253410, serial/lot #: (b)(6), udi#: (b)(4) ; product id: 8253410, serial/lot #: (b)(6), udi#: (b)(4).Product analysis: mainframe (s.No: (b)(6) unable to confirm the reported fault.The monitor cosmetically not ok.Line filter missing.Front bezel found broken.Imdrf codes: img g02030, fdd a0908, fdr c07, c070603, fdc d02 <(>&<)> fdm b01 product analysis: patient interface (s.No: (b)(6) the patient interface unit found cosmetically not okay.Patient interface cable faulty.Channel 3 not working.Fuses found blown and pcb found faulty.Imdrf codes: fdd a0908, img g02004, g02005 <(>&<)> g0201202 , fdd a0908, fdr c0208, fdc d02 <(>&<)> fdm b01.Product analysis: patient interface (s.No: (b)(6) the patient interface unit found cosmetically not okay.Patient interface cable faulty.Continuous beep coming after changing to mon itoring mode.Fuses found blown and pcb found faulty.Imdrf codes: fdd a0908, img g02004, g02005 <(>&<)> g0201202 , fdd a0908, fdr c0208, fdc d02 <(>&<)> fdm b01.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM-NEURO® 3.0 MAINFRAME
Common Device Name
STIMULATOR, NERVE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
glen belmer
6743 southpoint drive north
jacksonville, FL 32216
6122713209
MDR Report Key18089575
Report Number1045254-2023-00919
Device Sequence Number20521654
Product Code ETN
UDI-Device Identifier00613994529077
UDI-Public00613994529077
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K083124
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 11/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8253402
Device Catalogue Number8253402
Device Lot Number205963880
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/11/2023
Initial Report FDA Received Date11/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H10....".
Patient SexUnknown
-
-