| Model Number |
400CE |
| Medical Device Problem Code |
Insufficient Heating (1287)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complaint #(b)(4) received.There were no allegations of patient harm.The customer did not return the device for evaluation.
|
| |
|
Event or Problem Description
|
|
Customer reports: "we had an issue with one of our hemotherm (b)(4) units (s/n (b)(6), va biomed ee 194-11307ce) yesterday during our case.Upon rewarming the patient/unit, the warm side would not go above 34 degrees.We switched out units to continue rewarming our patient.There was not any alarm on the affected unit.".
|
| |
|
Additional Manufacturer Narrative
|
|
Udi related data quality updates only: udi information in section d of original submission is incorrect.
|
| |
|
Search Alerts/Recalls
|