Returned qty.(b)(4) inserts, 1.80395, (b)(6)-00093046, gg, beyond useful life test method / gp005-wi02 inductance: gauge id# 5349 due: 11/30/2024 result: 2.97 meets spec, does not meet spec, n/a.Tip condition: - meets spec, does not meet spec, n/a.Stack condition, meets spec, does not meet spec, n/a.Tested on: g136 gage id# 9626-2 due date: 03/31/24 set pressure: 35 psi.Water flow: output rate: (b)(4) - meets spec, does not meet spec, n/a.Spray pattern: - meets spec, does not meet spec, n/a.Water leak: hp sealing ring, proximal ring, distal ring, seam leak, n/a.Vibration: - meets spec, does not meet spec, n/a.Grip condition: meets spec, does not meet spec, n/a.Other visual observation: insert is good, no fault found.Also insert was tested on a digital thermometer i.D.#6116a due date: 09/30/2024.The temperature was 84.9° f, no fault found.The specification states are not to exceed 118.4°f.(per frs-9175).Alleged event: confirmed - not confirmed.2.80395, (b)(6)-00093046, gg, beyond useful life, test method / gp005-wi02.Inductance: gauge id# 5349 due: 11/30/2024 result: 3.13.Meets spec , does not meet spec, n/a.Tip condition: - meets spec, does not meet spec , n/a.Stack condition - meets spec, does not meet spec, n/a.Tested on: g136 gage id# 9626-2 due date: 03/31/24 set pressure: 35 psi.Water flow: output rate: 35 psi - meets spec, does not meet spec, n/a spray pattern: - meets spec, does not meet spec, n/a.Water leak: hp sealing ring, proximal ring, distal ring, seam leak, n/a.Vibration: - meets spec, does not meet spec, n/a., grip condition: meets spec, does not meet spec, n/a.Other visual observation: insert is good, no fault found.Also insert was tested on a digital thermometer i.D.#6116a due date: 09/30/2024.The temperature was 85.4° f, no fault found.The specification states are not to exceed 118.4°f.(per frs-9175).Alleged event: confirmed - not confirmed.3.80395, (b)(6)-00093046, gg, beyond useful life, test method / gp005-wi02.Inductance: gauge id# 5349, due: 11/30/2024, result: 3.11.Meets spec, does not meet spec, n/a, tip condition: - meets spec, does not meet spec, n/a.Stack condition , meets spec, does not meet spec, n/a.Tested on: g136 gage id# 9626-2 due date: 03/31/24 set pressure: 35 psi.Water flow: output rate: (b)(4)- meets spec, does not meet spec, n/a.Spray pattern: - meets spec, does not meet spec, n/a.Water leak: hp sealing ring, proximal ring , distal ring, seam leak, n/a.Vibration: - meets spec , does not meet spec, n/a.Grip condition: - meets spec , does not meet spec, n/a.Other visual observation: insert is good, no fault found.Also insert was tested on a digital thermometer i.D.#6116a due date: 09/30/2024.The temperature was 89.6° f, no fault found.The specification states are not to exceed 118.4°f.(per frs-9175).Alleged event: confirmed - not confirmed.
|