| Catalog Number |
165810 |
| Medical Device Problem Codes |
Incorrect Measurement (1383); Nonstandard Device (1420)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the foley catheters were flimsy and bending with resistance (pcn#165808 and pcn#165810).Additionally, the balloon seemed to be larger (pcn#165808 and pcn#165810) and was not deflating all the way resulting in tears and bleeding (pcn#165808 and pcn#165810).
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheters were flimsy and bending with resistance (pcn#165808 and pcn#165810).Additionally, the balloon seemed to be larger (pcn#165808 and pcn#165810) and was not deflating all the way resulting in tears and bleeding (pcn#165808 and pcn#165810).
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be due to "incorrect balloon design(inflation diameter undersized)".The dhr review could not be performed as lot provided in the complaint was invalid.A labeling review was not performed because labeling could not have prevented the reported failure.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the device was not returned.
|
| |
|