| Catalog Number |
Z304H |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Hold vm it was reported that a patient underwent an unknown procedure on an unknown date and suture was used.During the procedure, the needle was detached from the suture.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained via: "1st product : the suture could not be taken out of the tray.- please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.The device has been received at (b)(6) and will be shipped.Please check rmao.2nd product : the needle was detached from the suture.- when was the needle detached/pulled off from the suture (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify unk.- please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.The device has been received at (b)(6) and will be shipped.Please check rmao.3rd product : the needle was detached from the suture.(however, part of the needle remained on the suture.It may have been broken at the base.) - did the needle fall into the patient? no.If yes, ¿ was the needle piece(s) retrieved during the same procedure? ¿ were x-rays taken to locate the needle piece(s)? ¿ what measures were taken to retrieve the broken piece? ¿ was there any additional tissue damage as a result of searching for the needle piece? - does a piece of the needle remain in the patient¿s tissue? no.If yes, ¿ is there any plan in place to remove the needle piece in the future? ¿ if yes, please provide the scheduled date.¿ what tissue structure the broken needle was located? ¿ what is the surgeon's opinion of consequences to the patient? - please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.The device has been received at (b)(6) and will be shipped.Please check rmao.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Events reported via: 2210968-2023-08644.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 12/5/2023.H6 component code: g07002 no device problem found.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection evaluation was conducted on the returned device.Visual analysis of the returned sample determined that one detached needle and one suture piece outside its packaging that pertains to product code z304h were returned for analysis.During the inspection of the detached needle, the swage, and the attachment area were noted to be as expected.The barrel hole of the detached needle was examined under magnification, and no suture remnant was noted.The suture was examined, and a mark of correct insertion was noted in the extremes of the suture; however, the force used to detach the needle from the suture could not be determined.The functional test could not be performed due to the condition of the sample.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.No conclusion could be reached as to what caused the reported complaint.It should be noted that as part of our quality process, the device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|