| Model Number |
DM2102 |
| Medical Device Problem Codes |
Failure to Interrogate (1332); Incorrect, Inadequate or Imprecise Result or Readings (1535); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During an in-clinic follow-up, the device was unable to be interrogated and delivered an error message.In addition, previous episodes of undersensing were also observed on the device.Abbott technical support was contacted and suspected a corrupt egm was the cause of the event.No intervention was performed at this time.The patient was stable and will continue to be monitored.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated the patient was later seen in clinic, and the device was reprogrammed to resolve the event.The patient was in stable condition.
|
| |
|
Search Alerts/Recalls
|