ABBOTT VASCULAR DRAGONFLY¿ OPTIS¿ IMAGING CATHETER; DIAGNOSTIC INTRAVASCULAR CATHETER Back to Search Results
Catalog Number C408646
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/11/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event or Problem Description
It was reported that during the procedure, the dragonfly optis imaging catheter was leaking contrast from a hole at the tip of the device while being used in the patient.The device was removed, and a new dragonfly device was used successfully for the procedure.There were no reported adverse patient effects and no clinically significant delays in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Visual analysis and functional testing were performed on the returned device.The reported leak and ¿a hole at the tip of the device¿ were unable to be confirmed, as the returned device was able to be purged as expected with purged liquid exiting the designated purge hole.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.A query of the complaint handling database for the reported lot was performed and revealed there is no indication of a lot level quality product issue.The investigation determined that the cause for reported leak and ¿a hole at the tip of the device¿ were unable to be confirmed.The catheter was returned with dried contrast visible throughout the entirety of the sheath, so the catheter was placed into a water bath in attempt to remove the dried contrast.After the water bath, purging liquid through the catheter was attempted again and the purged liquid was able to flow freely through the sheath, as expected, and exited through the purge hole.There was no damage/hole noted to the catheter distal tip, and liquid exited proximally from the distal tip and through the dedicated purge hole.It should be noted that optis and opstar model catheters have a dedicated purge hole located proximal to the guidewire exchange minirail, on the distal portion of the sheath.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRAGONFLY¿ OPTIS¿ IMAGING CATHETER
Common Device Name
DIAGNOSTIC INTRAVASCULAR CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT MEDICAL REG#3009600098
4 robbins rd
westford MA 01886
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18093368
Report Number2024168-2023-12406
Device Sequence Number5945216
Product Code DQO
UDI-Device Identifier00183739000654
UDI-Public00183739000654
Combination Product (Y/N)N
Initial Reporter CountryTC
PMA/510(K) Number
K141769 
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/09/2025
Device Catalogue NumberC408646
Device Lot Number8867967
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/18/2023
Supplement Date Received by Manufacturer01/26/2024
Initial Report FDA Received Date11/08/2023
Supplement Report FDA Received Date02/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-