STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INSERT #5 16MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
| Catalog Number |
5531-G-516-E |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
10/12/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding disassociation involving a triathlon insert was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, lot number of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.H3 other text : device not available.
|
| |
|
Event or Problem Description
|
|
It was reported by the patient that on (b)(6) 2023, she underwent a mako assisted robotic surgery for a right total knee replacement.Allegedly x-ray taken immediately post op showed the locking mechanism was in an upright position whereas it should have been locked down until it could not be seen at all.The patient underwent a revision surgery later the same day to have the insert replaced.
|
| |
|
Search Alerts/Recalls
|
|
|