• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012449-15
Medical Device Problem Codes Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/03/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported difficulty advancing and removing the device from the guide wire.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.The ht bmw universal ii referenced in b5 is filed under a separate medwatch report number.
 
Event or Problem Description
It was reported that the procedure was to treat a stenosed lesion in the left anterior descending (lad) artery.The hi-torque (ht) balance middleweight (bmw) universal ii guide wire was advanced and used as a track.A stent was implanted and a non-compliant balloon was used for post dilatation.The 3.0x15mm nc trek balloon dilatation catheter (bdc) was attempted to be advanced to the stent but with significant resistance with the guide wire, could not reach the site where it needed to be dilated.A 6french (fr) guiding catheter also failed to pass into the artery.The ht bmw universal ii guide wire and the nc trek bdc were removed together as a unit.Another non-abbott guide wire and a smaller size non-abbott balloon were used to complete the procedure.However, dilatation of the stent may not be complete which may lead to in-stent restenosis.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18096632
Report Number2024168-2023-12434
Device Sequence Number17496245
Product Code LOX
UDI-Device Identifier08717648151873
UDI-Public08717648151873
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2026
Device Catalogue Number1012449-15
Device Lot Number30503G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/19/2023
Initial Report FDA Received Date11/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6FR GUIDE CATHETER.; HT BMW GUIDE WIRE.
Patient SexUnknown
-
-