• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH DOUBLE AIR HOSE, 3M, FOR AIR PEN DRIVE; MOTOR, DRILL, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH DOUBLE AIR HOSE, 3M, FOR AIR PEN DRIVE; MOTOR, DRILL, PNEUMATIC Back to Search Results
Catalog Number 05.001.083
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.The double air hose device was evaluated and the reported condition that the inner hose came off was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the inner hose ruptured.The assignable root cause of this condition was determined to be traced to component failure due to wear.Udi - (b)(4).
 
Event or Problem Description
It was reported from switzerland that during service and evaluation, it was determined that the double air hose device ruptured.It was noted that the inner hose tore off.It was further determined that the device failed pretest for internal pressure test.It was noted in the service order that the inner hose had come off.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOUBLE AIR HOSE, 3M, FOR AIR PEN DRIVE
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key18096902
Report Number8030965-2023-14125
Device Sequence Number5709428
Product Code HBB
UDI-Device Identifier07611819831520
UDI-Public07611819831520
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K093361
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number05.001.083
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2023
Initial Report FDA Received Date11/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/16/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-