| Catalog Number |
PGCC300 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "product bent during insertion which made it unusable".No patient harm or injury.The patient status is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that "product bent during insertion which made it unusable".No patient harm or injury.The patient status is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The complaint sample was not returned.Grace manufacturing, the manufacturing site reported: "no sample was available for testing.The dhr was reviewed.All operations and documentation were completed.The cause of the damaged device is undetermined.It is possible the device was mishandled and misused by the user resulting in a bent jaw.The root cause is unknown, and no corrective actions are necessary currently." teleflex will continue to monitor and trend for reports of this nature.
|
| |
|