| Model Number |
M0062101400 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0402 captures the reportable event of balloon burst.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a nephromax nephrostomy balloon catheter device was used during a ureteroscopy (urs) procedure performed on (b)(6) 2023.During the procedure, the balloon pop when rising the pressure to 12 atm (atmosphere).The procedure was completed with another nephromax nephrostomy balloon catheter device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be good.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h2: additional information: block d4 (lot number) have been updated based on the additional information received on november 9, 2023.Block h6: imdrf device code a0402 captures the reportable event of balloon burst.Block h10: investigation result.A visual examination of the returned device found a longitudinal tear on the balloon measuring approximately 25mm in length located in the proximal region of the balloon.A visual and tactile examination was performed to the tip, catheter shaft and the markerband, and no damages were found.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.As this device is recommended for dilatation of the nephrostomy tract, and was used during a ureteroscopy procedure, it is possible that the balloon had an interaction with a stone which would have contributed to the rupture in the balloon material.Therefore, the most probable root cause is adverse event related to procedure.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a nephromax nephrostomy balloon catheter device was used during a ureteroscopy (urs) procedure performed on (b)(6) 2023.During the procedure, the balloon pop when rising the pressure to 12 atm (atmosphere).The procedure was completed with another nephromax nephrostomy balloon catheter device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be good.
|
| |
|