• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NEPHROMAX; CATHETER, NEPHROSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NEPHROMAX; CATHETER, NEPHROSTOMY Back to Search Results
Model Number M0062101400
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a0402 captures the reportable event of balloon burst.
 
Event or Problem Description
It was reported to boston scientific corporation that a nephromax nephrostomy balloon catheter device was used during a ureteroscopy (urs) procedure performed on (b)(6) 2023.During the procedure, the balloon pop when rising the pressure to 12 atm (atmosphere).The procedure was completed with another nephromax nephrostomy balloon catheter device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be good.
 
Additional Manufacturer Narrative
Block h2: additional information: block d4 (lot number) have been updated based on the additional information received on november 9, 2023.Block h6: imdrf device code a0402 captures the reportable event of balloon burst.Block h10: investigation result.A visual examination of the returned device found a longitudinal tear on the balloon measuring approximately 25mm in length located in the proximal region of the balloon.A visual and tactile examination was performed to the tip, catheter shaft and the markerband, and no damages were found.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.As this device is recommended for dilatation of the nephrostomy tract, and was used during a ureteroscopy procedure, it is possible that the balloon had an interaction with a stone which would have contributed to the rupture in the balloon material.Therefore, the most probable root cause is adverse event related to procedure.
 
Event or Problem Description
It was reported to boston scientific corporation that a nephromax nephrostomy balloon catheter device was used during a ureteroscopy (urs) procedure performed on (b)(6) 2023.During the procedure, the balloon pop when rising the pressure to 12 atm (atmosphere).The procedure was completed with another nephromax nephrostomy balloon catheter device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEPHROMAX
Common Device Name
CATHETER, NEPHROSTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
5086834015
MDR Report Key18097849
Report Number2124215-2023-62038
Device Sequence Number12263084
Product Code LJE
UDI-Device Identifier08714729826484
UDI-Public08714729826484
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K121614
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/24/2025
Device Model NumberM0062101400
Device Catalogue Number210-140
Device Lot Number0030412928
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/20/2023
Supplement Date Received by Manufacturer11/09/2023
Initial Report FDA Received Date11/08/2023
Supplement Report FDA Received Date11/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-