| Catalog Number |
46104-65 |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation; however, testing has not yet been completed.
|
| |
|
Event or Problem Description
|
|
The event involved a the event involved a tp4 kit w/10cc safeset reservoir, needleless valve, filter millex and admin set.It was reported that when hanging aline tubing, it was connected to patient and was not able to flush line, and blood started to backflow into the line.There was patient involvement and no patient harm reported.The set-up was replaced and therapy was resumed.There were no interventions needed related to blood loss.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported complaint of not able to flush was confirmed on the returned set.An image was provided by the customer showing the involved device.No visual anomalies were observed on the returned device.When the returned device was primed and pressure leak tested the set was not able to be primed from the millex filter.The rest of the set performed per the specifications.When the millex filter was microscopically examined, the female luer connected to the millex filter was occluded with solvent.The probable cause of the occlusion on the female luer had occurred due to an error during assembly at ensenada.The lot history was reviewed and no nonconformities were identified that may have contributed to the reported complaint.
|
| |
|