This device is classified as import for export, therefore 510k is not applicable.Model fb-18v is available in the usa with a 510k number k951199.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the cfb (coherent fiber bundle) broken.Based on the result, we concluded that it was caused due to the excessive force applied on the cfb (coherent fiber bundle).In addition, our technician confirmed that the control body fluid damage, the forward body frame fluid damage, the ocular fluid damage, the cfb screen cover glass fluid damage, the screen mask fluid damage, the side body cover broken, the insertion flexible tube perforated, the operation channel (primary) crushed, the insertion flexible tube crushed, and the segment crushed; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|