| Model Number |
DRILL, 2.75MM, .066 CANNULATION |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
10/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 10/16/2023, it was reported by a sales representative via email that an ar-7000-14 2.75 mm drill broke off inside the patient right after the first ar-7000-26ft cannulated screw, fully threaded was placed.A c-arm was brought in to locate the broken fragment.Once the fragment was found, the surgeon was unable to retrieve it.The surgeon retained the broken fragments using another ar-7000-26ft cannulated screw, fully threaded, which was manually implanted since there was no other drill available to implant the rest of the screw.This was discovered during a distal tibial allograft placement on the shoulder on (b)(6) 2023.The case was delayed for about thirty minutes.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint allegation is confirmed.One unpackaged ar-7000-14 serial/batch number (b)(6) was received for evaluation.Functional testing was not performed due to damage to the device.Visual inspection found that the device was broken at the proximal and distal ends.The drill's overall length was measured by drawing specification c1455 at revision 5 (9.00 ±.01).Results.Due to the breakage, the device is shorter, measuring 7.20 in.The observed condition is likely due to prying/levering the device against hard bone or other instrumentation during use.
|
| |
|
Search Alerts/Recalls
|