| Model Number |
HEADED REAMER, 7.5 MM CANNULATED |
| Medical Device Problem Code |
Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that ar-1407.5, sn: (b)(6) is dull.There was no harm for patient, operator or third party reported.No further information received.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 complaint is confirmed.Visual evaluation identified around the edges of the cutting flutes of the returned cannulated headed reamer, 7.5 mm, had dulled.Abrasion marks and chips around the flutes were noted.Functional testing was not performed due to the damage to the device.The most likely cause of this failure is attributed to wear and tear damage incurred over repeated usage and manufactured in 2012.
|
| |
|