• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC PEDIATRIC CRANIOTOME; MOTOR, DRILL, ELECTRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC PEDIATRIC CRANIOTOME; MOTOR, DRILL, ELECTRIC Back to Search Results
Catalog Number CRANI-P-G1
Medical Device Problem Codes Device Damaged by Another Device (2915); Complete Loss of Power (4015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.The device was evaluated and it was determined that the reported condition was not confirmed.Therefore, the assignable root cause was not determined.However, during evaluation, it was determined that the device tip was bent.It was further determined that the device failed pretest for visual assessment.The assignable root cause was traced to component failure due to normal wear.Udi: (b)(4).
 
Event or Problem Description
It was reported from switzerland that during service and evaluation, it was determined that the craniotome device tip was bent.It was further determined that the device failed pretest for visual assessment.It was noted in the service order that the device did not work prior to a surgical procedure.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEDIATRIC CRANIOTOME
Common Device Name
MOTOR, DRILL, ELECTRIC
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
3035526892
MDR Report Key18102143
Report Number1045834-2023-01589
Device Sequence Number20258276
Product Code HBC
UDI-Device Identifier00845384016403
UDI-Public00845384016403
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K133604
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCRANI-P-G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2023
Initial Report FDA Received Date11/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/04/2015
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-