| Model Number |
LSKB1-BX-05 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
10/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The removal of the anchor bolt was confirmed by extra x-rays.
|
| |
|
Event or Problem Description
|
|
Upon the removal of the head wrap the surgeon noted that an anchor bolt was broken.Additional medical intervention was needed to remove the bolt; however there was no harm or injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The removal of the anchor bolt was confirmed by extra x-rays.Updated 12/6/2023.The product was returned for evaluation.The alleged complaint was confirmed that the anchor bolt was broken.The anchor bolt broke at the point it was designed to break.Breaking can occur from many things like excessive force when tightening, patient movement (seizures) during monitoring, etc.It is unclear at what point in time the anchor bolt broke as it was not discovered until the removal of the head wrap.No root cause can be positively identified of the return or the information supplied.The calculated occurrence level matches the occurrence level present in the risk file and the resulting risk level will remain "alap".No further action is needed at this time.
|
| |
|
Event or Problem Description
|
|
Upon the removal of the head wrap the surgeon noted that an anchor bolt was broken.Additional medical intervention was needed to remove the bolt; however there was no harm or injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The removal of the anchor bolt was confirmed by extra x-rays.Updated 12/6/2023.The product was returned for evaluation.The alleged complaint was confirmed that the anchor bolt was broken.The anchor bolt broke at the point it was designed to break.Breaking can occur from many things like excessive force when tightening, patient movement (seizures) during monitoring, etc.It is unclear at what point in time the anchor bolt broke as it was not discovered until the removal of the head wrap.No root cause can be positively identified of the return or the information supplied.The calculated occurrence level matches the occurrence level present in the risk file and the resulting risk level will remain "alap".No further action is needed at this time."udi related data quality updates only." updated 09/20/2024.Clarification and correction of the following missing items from previous supplemental report: combination product: no.Brand name: anchor bolt kit, lsb style.Model and catalog number: lskb1-bx-05.Primary di number: (b)(4).
|
| |
|
Event or Problem Description
|
|
Upon the removal of the head wrap the surgeon noted that an anchor bolt was broken.Additional medical intervention was needed to remove the bolt; however, there was no harm or injury to the patient.
|
| |
|