• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER; MIXER, BREATHING GASES, ANESTHESIA INHALATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER; MIXER, BREATHING GASES, ANESTHESIA INHALATION Back to Search Results
Model Number MICBLNDR,W/O HOSES,DOM,H/L
Medical Device Problem Code Inaccurate Delivery (2339)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/13/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: 81 other - at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.H3 other text : device not returned yet.
 
Event or Problem Description
It was reported to vyaire medical that the micro blender is not sending out the correct o2 (oxygen) percentage to patients.The customer confirmed that there is no patient harm associated with this reported event.
 
Additional Manufacturer Narrative
Result of investigation: vyaire failure analysis was duplicate the reported issue.Blender was found to function properly however deliver o2 (oxygen) inaccurately.Depending on gas line conditions, possible contaminants may compromise the function of the micro blender causing the unit to not meet its specified performance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER
Common Device Name
MIXER, BREATHING GASES, ANESTHESIA INHALATION
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
sandra valencia
510 technology drive
irvine, CA 92618
8477865998
MDR Report Key18102650
Report Number2021710-2023-18383
Device Sequence Number18115600
Product Code BZR
UDI-Device Identifier10846446002348
UDI-Public(01)10846446002348
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K883038
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMICBLNDR,W/O HOSES,DOM,H/L
Device Catalogue Number03800A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2023
Supplement Date Received by Manufacturer12/01/2023
Initial Report FDA Received Date11/09/2023
Supplement Report FDA Received Date12/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-