| Model Number |
MICBLNDR,W/O HOSES,DOM,H/L |
| Medical Device Problem Code |
Inaccurate Delivery (2339)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: 81 other - at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.H3 other text : device not returned yet.
|
| |
|
Event or Problem Description
|
|
It was reported to vyaire medical that the micro blender is not sending out the correct o2 (oxygen) percentage to patients.The customer confirmed that there is no patient harm associated with this reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Result of investigation: vyaire failure analysis was duplicate the reported issue.Blender was found to function properly however deliver o2 (oxygen) inaccurately.Depending on gas line conditions, possible contaminants may compromise the function of the micro blender causing the unit to not meet its specified performance.
|
| |
|