MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 WITH SHUNTASSISTANT 20; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 WITH SHUNTASSISTANT 20; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX413T
Medical Device Problem Codes Mechanical Problem (1384); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 10/16/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation on-going.Shoud relevant additional iinfornation/investigation results became available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported that a progav 2.0 shunt system (#fx413t) was implanted during a procedure performed on (b)(6) 2022.According to the complainant, the shunt showed a over drainage and adjustment problems.The patient underwent a revision procedure performed on (b)(6) 2023.The complainant device has not yet returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 12 months.Weight: 8.2 kilograms.Height: 72 centimeters.Gender: female.
 
Additional Manufacturer Narrative
Investigation: visual inspection during the investigation, no significant deformations or damage of the valves was determined.Permeability test a permeability test has shown that all components are permeable.The drained liquid was clear.Computer controlled test to investigate the claim of over- drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the progav 2.0 is operating not within the accepted tolerance in the horizontal position.The shuntassistant is operating within the specified tolerances in both the horizontal as well as the vertical positions.An accelerated outflow of progav 2.0 could be determined.Adjustment test the progav 2.0 was tested and is not adjustable.Braking force and brake function test the brake functionality test has shown that the brake function is operational, however the braking force cannot be measured due to the non-adjustability of the valve.Internal inspection after dismantling of the valves, deposits were found in both.To make the proteins / deposits in the shunt system more visible, they were colored using a staining solution.Results based on our investigation results, we can determine an accelerated outflow and a non-adjustability at the progav 2.0.The visible deposits in the valve may have led to the functional impairment.Deposits caused by substances naturally present in the patient's bodies, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic material can affect the integrity of the valve.Also, we can determine visible deposits at the shuntassistant.The deposits had no effect of the valve upon the specifications at the time of the examination.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 WITH SHUNTASSISTANT 20
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key18104101
Report Number3004721439-2023-00328
Device Sequence Number12151562
Product Code JXG
UDI-Device Identifier04041906136550
UDI-Public4041906136550
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/08/2026
Device Model NumberFX413T
Device Catalogue NumberFX413T
Device Lot Number20054564
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2023
Supplement Date Received by Manufacturer11/14/2023
Initial Report FDA Received Date11/09/2023
Supplement Report FDA Received Date12/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age12 MO
Patient SexFemale
Patient Weight8 KG
-
-