C.R. BARD, INC. (COVINGTON) -1018233 BARD® DUAL LUMEN URETERAL CATHETER Back to Search Results
Catalog Number 130200
Medical Device Problem Codes Material Fragmentation (1261); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/16/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that tip of ureteral catheter was smashed down or defective.Staff were not able to insert guidewire as they usually can.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that tip of ureteral catheter was smashed down or defective.Staff were not able to insert guidewire as they usually can.
 
Additional Manufacturer Narrative
The reported event was inconclusive since no sample was returned for evaluation.A potential root cause for this failure mode could be ¿incorrect geometry of tip".The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "directions for use: the ureteral catheter should be inserted into the ureter using standard endourologic technique, under direct, fluoroscopic or radiographic visualization.Caution: avoid sharp bending.This is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and with applicable local, state and federal laws and regulations." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® DUAL LUMEN URETERAL CATHETER
Common Device Name
URETERAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key18104240
Report Number1018233-2023-08003
Device Sequence Number12119767
Product Code EYB
UDI-Device Identifier00801741052903
UDI-Public(01)00801741052903
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K032521
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number130200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/16/2023
Supplement Date Received by Manufacturer03/01/2024
Initial Report FDA Received Date11/09/2023
Supplement Report FDA Received Date03/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-