Catalog Number 9-PDAP-05-04-L
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Codes Dyspnea (1816); Obstruction/Occlusion (2422)
Date of Event 10/18/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
The article, "left pulmonary artery occlusion following device closure of patent ductus arteriosus in premature infants", was reviewed.The article presented a case study of a 25-day-old, 975g, male patient with comorbidities persistent metabolic acidosis, ventilator dependence, twin died due to necrotizing enterocolitis (nec) with large patent ductus arteriosus (pda), bronchopulmonary dysplasia (bpd), and poor nutrition.It was reported on an unknown date a 5-4mm amplatzer piccolo occluder was chosen for implant to occluder a pda with the following dimensions: narrowest diameter = 3.3mm, largest diameter = 4.3mm, and length = 12mm.It was reported 12 weeks post-procedure, transthoracic echocardiography showed an increase in maximal gradient velocity with minimal flow by color doppler.The patient additionally required respiratory support.An angiography revealed severe obstruction to the lpa.A decision was made to attempt percutaneous revascularization of the lpa but unsuccessful.The patient was then admitted for surgical revascularization of the lpa and the proximal disc of the 5-4mm amplatzer piccolo occluder was excised to relieve the obstruction.The patient status was reported discharged with excellent lpa flow and only mild turbulence at 3 month follow up.The article concluded although rare, severe lpa obstruction can be seen in premature infants following device closure of the pda.The piccolo device is designed to ideally remain entirely intraductal.Although device selection appeared to provide a device short enough for the given ductal length, they recommend, whenever possible, giving consideration to using the shortest possible device.We also recommend increasing the frequency of echocardiographic surveillance to weekly studies if at any time the imaging demonstrates an increase in the degree of obstruction/turbulence.Clinicians should have a low threshold for alternative imaging modalities such as computed tomography or more invasive options if a pattern of increased obstruction is detected.This may allow for a transcatheter based intervention to either retrieve the device, if relatively soon, or stent the lpa, if patient size allows.[the primary and corresponding author was jesus jaile, children¿s hospital of new orleans / lsuhsc, new orleans, louisiana, with corresponding email: jesus.Jaile@lcmchealth.Org].
 
Additional Manufacturer Narrative
Summarized patient outcomes/complications of amplatzer piccolo were reported in a research article in a subject population with multiple co-morbidities including persistent metabolic acidosis, ventilator dependence, twin died due to necrotizing enterocolitis (nec) with large patent ductus arteriosus (pda), bronchopulmonary dysplasia (bpd), and poor nutrition.Some of the complications reported were unexpected medical intervention, surgical intervention, hospitalization, obstruction/occlusion, and dyspnea.; these complications are anticipated for the procedure and subject population.A more comprehensive assessment could not be performed as the event was non-contemporaneously reported through a literature review and no device or individual patient information was received for analysis.  per the instructions for use, it stated, " warnings: do not release the occluder from the delivery wire if either a retention disc protrudes into the pulmonary artery or aorta; or if the position of the occluder is not stable." this is considered as an improper or incorrect procedure or method of the device.However, it was unable to determine if an improper or incorrect procedure or method contributed to the reported incident.Therefore, a letter will not be provided to address the deviation from the ifu.
 
Event or Problem Description
The article, "left pulmonary artery occlusion following device closure of patent ductus arteriosus in premature infants", was reviewed.The article presented a case study of a 25-day-old, 975g, male patient with comorbidities persistent metabolic acidosis, ventilator dependence, twin died due to necrotizing enterocolitis (nec) with large patent ductus arteriosus (pda), bronchopulmonary dysplasia (bpd), and poor nutrition.It was reported on an unknown date a 5-4mm amplatzer piccolo occluder was chosen for implant to occluder a pda with the following dimensions: narrowest diameter = 3.3mm, largest diameter = 4.3mm, and length = 12mm.During procedure, it was reported device implant was straightforward, and post release transthoracic echocardiography (tte) showed it to be well seated with only slight left pulmonary artery (lpa) turbulence caused by the proximal disc.It was then reported 12 weeks post-procedure, tte showed an increase in maximal gradient velocity with minimal flow by color doppler.The patient additionally required respiratory support.An angiography revealed severe obstruction to the lpa caused by the proximal disc.A decision was made to attempt percutaneous revascularization of the lpa but unsuccessful.The patient was then admitted for surgical revascularization of the lpa and the proximal disc of the 5-4mm amplatzer piccolo occluder was excised to relieve the obstruction.The patient status was reported discharged with excellent lpa flow and only mild turbulence at 3 month follow up.The article concluded although rare, severe lpa obstruction can be seen in premature infants following device closure of the pda.The piccolo device is designed to ideally remain entirely intraductal.Although device selection appeared to provide a device short enough for the given ductal length, they recommend, whenever possible, giving consideration to using the shortest possible device.We also recommend increasing the frequency of echocardiographic surveillance to weekly studies if at any time the imaging demonstrates an increase in the degree of obstruction/turbulence.Clinicians should have a low threshold for alternative imaging modalities such as computed tomography or more invasive options if a pattern of increased obstruction is detected.This may allow for a transcatheter based intervention to either retrieve the device, if relatively soon, or stent the lpa, if patient size allows.[the primary and corresponding author was (b)(6), with corresponding email: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER PICCOLO
Common Device Name
OCCLUDER, PATENT DUCTUS, ARTERIOSUS
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18104423
Report Number2135147-2023-04947
Device Sequence Number5556127
Product Code MAE
UDI-Device Identifier05415067031044
UDI-Public05415067031044
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
P020024
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Literature,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number9-PDAP-05-04-L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/19/2023
Supplement Date Received by Manufacturer12/13/2023
Initial Report FDA Received Date11/09/2023
Supplement Report FDA Received Date01/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age25 DA
Patient SexMale
-
-