| Catalog Number |
07251068190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer's e 801 analyzer serial number is (b)(6).The serial number of the e 801 analyzer used for investigation was requested, but not provided.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received questionable results for one patient sample tested with elecsys pth on two cobas e 801 module analyzers.Refer to the attachment for all patient data.The sample was initially tested on the customer's e 801 analyzer on (b)(6) 2023.The customer repeated the sample with x2, x4, and x8 dilutions.The patient's previous pth values ranged between 50 and 100 pg/ml, so the results from the complained sample were questioned.The sample was provided for investigation where it was tested with pth and pth (1-84) on a second e 801 analyzer on (b)(6) 2023.Pth and pth (1-84) were also re-measured after x2, x4, and x8 manual dilutions with both low pth patient serum and diluent solution.
|
| |
|
Additional Manufacturer Narrative
|
|
Calibration and control data were acceptable.There was no indication of a reagent issue.The sample was not available for further investigation.The investigation could not identify a product problem.The cause of the event could not be determined.
|
| |
|