• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES C4130, 5MMX35CM EPIX LAPA GRASPER, 10/BX; LAPAROSCOPIC ACCESSORIES, GYNECOLOGIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES C4130, 5MMX35CM EPIX LAPA GRASPER, 10/BX; LAPAROSCOPIC ACCESSORIES, GYNECOLOGIC Back to Search Results
Model Number C4130
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/12/2023
Type of Reportable Event Malfunction
Event or Problem Description
Name of procedure being performed: na (before use) detailed description of event 13-oct-2023 - complaint product family: grasper, product name: epix grasper, model number: c4130, lot number: 1468046.- the event occurred before use of the device, for an unknown procedure.- the nurse had identified whilst doing set-up.It was reported that hair had been found inside the sealed package of the device.- there was no clinical impact to the patient as a result of before use of the procedure.- the user resolved the situation by using another device from the shelf.- no other instruments were used when the complaint event occurred.Patient status: na (no patient involvement) type of intervention: na (before use).
 
Additional Manufacturer Narrative
The event device is anticipated to return.A follow-up report will be provided upon completion of the investigation.
 
Event or Problem Description
Name of procedure being performed: na (before use).Detailed description of event 13-oct-2023.Complaint product family: grasper, product name: epix grasper, model number: c4130, lot number: 1468046.The event occurred before use of the device, for an unknown procedure.The nurse had identified whilst doing set-up.It was reported that hair had been found inside the sealed package of the device.There was no clinical impact to the patient as a result of before use of the procedure.The user resolved the situation by using another device from the shelf.No other instruments were used when the complaint event occurred.Patient status: na (no patient involvement).Type of intervention: na (before use).
 
Additional Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the presence of hair inside the sterile unopened pouch.Based on the condition of the returned unit and the description of the event, the hair likely originated from an operator during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C4130, 5MMX35CM EPIX LAPA GRASPER, 10/BX
Common Device Name
LAPAROSCOPIC ACCESSORIES, GYNECOLOGIC
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
farah azmi
22872 avenida empresa
rancho santa margarita, CA 92688
9497138710
MDR Report Key18107291
Report Number2027111-2023-00713
Device Sequence Number5698611
Product Code NWV
UDI-Device Identifier00607915110147
UDI-Public(01)00607915110147(17)251103(30)01(10)1468046
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/03/2025
Device Model NumberC4130
Device Catalogue Number101419801
Device Lot Number1468046
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2023
Supplement Date Received by Manufacturer10/13/2023
Initial Report FDA Received Date11/09/2023
Supplement Report FDA Received Date01/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-