• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC CAPSTONE® SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC CAPSTONE® SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, Back to Search Results
Model Number 2990003
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from a healthcare professional via a manufacturer representative regarding a patient with preoperative diagnosis for lumbar spinal stenosis.Procedure or therapy performed was l5/s mis -tlif.It was reported that the cage backed out after surgery.Gripping by shaft was done during cage removal.The gripping portion at the tip of the shaft was broken due to the twisting motion during removal.It was buried in the cage.Since the rear portion of the cage had already emerged, removal was performed with needle-nose pliers.Due to the narrow operative field, an external tube was not used.There was no patient symptoms or complications as a result of this event.Date of initial surgery was (b)(6) 2022.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Product analysis product id: 2990003; lot no: nm20e030 visual inspection confirmed the entire threaded tip of instrument has been sheared off.Macroscopic examination of the fracture surface revealed a brittle fracture with no indication of fatigue.This type of damage is consistent with bend stress overload.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSTONE® SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key18108658
Report Number1030489-2023-00742
Device Sequence Number20739232
Product Code MAX
UDI-Device Identifier00613994846266
UDI-Public00613994846266
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103731
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2990003
Device Catalogue Number2990003
Device Lot NumberNM20E030
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2023
Supplement Date Received by Manufacturer04/04/2024
Initial Report FDA Received Date11/10/2023
Supplement Report FDA Received Date04/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/10/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1