MAUDE Adverse Event Report: COLOPLAST A/S FOLYSIL SILICONE CATHETER; INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S FOLYSIL SILICONE CATHETER; INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL Back to Search Results
Catalog Number AA6112
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2023
Type of Reportable Event Malfunction
Event or Problem Description
According to the available information, after fitting the bladder catheter, it was impossible to instill the chemotherapy (gemcitabine 1g/50ml, 6th instillation).The bladder catheter was removed, and a permeability test carried out; the catheter was not permeable.A new catheter from the same batch was inserted after permeability had been checked.
 
Additional Manufacturer Narrative
After receiving this complaint, we searched for other complaint and we didn¿t find other complaint on the lot number 8931413.In november, we received one used sample.After desinfection, drainage was tested and conformed, inflation/deflation functionality was tested and conformed.No issue was observed about this catheter, so no root cause can be identified.A similar case study based on same item number (b)(4) same defect permeability issue over last four year, no similar case was found.Checking the quality database revealed no anomaly in relation with the described product.
 
Event or Problem Description
According to the available information, after fitting the bladder catheter, it was impossible to instill the chemotherapy (gemcitabine 1g/50ml, 6th instillation).The bladder catheter was removed, and a permeability test carried out; the catheter was not permeable.A new catheter from the same batch was inserted after permeability had been checked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLYSIL SILICONE CATHETER
Common Device Name
INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer (Section G)
CMF-SARLAT
9 avenue edmond rostand
sarlat-la-caneda
FR  
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key18110494
Report Number9610711-2023-00241
Device Sequence Number7717882
Product Code EZL
UDI-Device Identifier03600040127230
UDI-Public3600040127230
Combination Product (Y/N)N
PMA/510(K) Number
K013174
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberAA6112
Device Lot Number8931413_AA61121002
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/10/2023
Supplement Report FDA Received Date05/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-