| Catalog Number |
1012630-59 |
| Medical Device Problem Codes |
Failure to Advance (2524); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported difficulties and treatment appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat the subclavian artery with heavy calcification and no tortuosity.The 7.0x59mm omnilink elite stent delivery system (sds) had resistance with the anatomy during advancement and the stent dislodged from the balloon.Surgery was performed to remove the dislodged stent and nothing remains in the patient.Another same size omnilink elite stent was used to complete the procedure.There was no adverse patient sequalae and there was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|