| Catalog Number |
ES2814 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
According to the available information the device was explanted and replaced due to a malfunction of the prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, non-conforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed in veeva to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information the device was explanted and replaced due to a malfunction of the prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
Titan touch pump, cylinders 1 and 2, and reservoir were received for evaluation.Abrasion was noted on both exhaust tubes and the inlet tube of the pump.No functional abnormalities were noted with the pump.A separation, surrounded by abrasion, was noted on the exhaust tubing of cylinder 1, at the strain relief junction.No functional abnormalities were noted with cylinder 2.No functional abnormalities were noted with the reservoir.Based on examination of the returned product, it was concluded that the abrasion marks noted on both exhaust tubes and inlet tube of the pump matched in such a way to conclude that they had overlapped and abraded against one another while in-vivo enough to cause a separation in the exhaust tubing of cylinder 1 at the strain relief junction.A separation of this type could then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
|
| |
|
Event or Problem Description
|
|
According to the available information the device was explanted and replaced due to a malfunction of the prosthesis.
|
| |
|