• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. LIVEWIRE¿ TC ABLATION CATHETER 4 MM TIP UNIVERSAL TEMPERATURE MONITORING QUADRIP; Cardiac ablation percutaneous catheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. LIVEWIRE¿ TC ABLATION CATHETER 4 MM TIP UNIVERSAL TEMPERATURE MONITORING QUADRIP; Cardiac ablation percutaneous catheter Back to Search Results
Model Number 402152
Medical Device Problem Code Electrical Shorting (2926)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
This report is to advise of a short circuit that was noted during analysis.
 
Additional Manufacturer Narrative
One 7f, quadripolar, large curl, xls livewire tc ablation catheter was received for evaluation.A short circuit was detected between electrode 1 and conductor wire 2.Dissection revealed that the flat wire insulation was torn and the insulation for conductor wire 2 was abraded in the same location, consistent with the short circuit detected and the reported signal issue.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn insulation and abraded wires remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIVEWIRE¿ TC ABLATION CATHETER 4 MM TIP UNIVERSAL TEMPERATURE MONITORING QUADRIP
Common Device Name
Cardiac ablation percutaneous catheter
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
14901 deveau place
minnetonka MN 55345
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
14901 deveau place
minnetonka MN 55345
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key18115040
Report Number2182269-2023-00050
Device Sequence Number17896569
Product Code LPB
UDI-Device Identifier05414734200257
UDI-Public05414734200257
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
P960016
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2025
Device Model Number402152
Device Lot Number8356554
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2023
Initial Report FDA Received Date11/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-