• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH TARGETING ARM FEMUR GT T2 ALPHA FEMUR ANTEGRADE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH TARGETING ARM FEMUR GT T2 ALPHA FEMUR ANTEGRADE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 23533103
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code Perforation (2001)
Date of Event 10/14/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.A review of the device history is not possible because the lot number was not communicated.Should additional information become available, it will be provided in a supplemental report.
 
Event or Problem Description
As reported: "during implantation of alpha greater trochanter nail, the wire for lag screw placement entered the lateral wall of the femur but then missed the nail and ended anterior of the nail.This caused a delay of 20 min in the case as multiple attempts were needed to correctly place the wire.No replacements were necessary.".
 
Event or Problem Description
As reported: "during implantation of alpha greater trochanter nail, the wire for lag screw placement entered the lateral wall of the femur but then missed the nail and ended anterior of the nail.This caused a delay of 20 min in the case as multiple attempts were needed to correctly place the wire.No replacements were necessary.".
 
Additional Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidence was provided.Reasons for mal alignment between wire and nail¿s drill holes are various, e.G.[but not limited to] insufficient tightening of the nail holding screw, applying too much [axial] force on the construct or unsuitable bone geometry.Functional check prior to use is required in the labelling ¿ such as ifu, op-tech and re-processing brochure.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling was not possible because the catalog number and lot number were not communicated.Indications of material, manufacturing, or design related problems were unable to be identified as the catalog number and lot number were not communicated.In case substantive information will become available in future that suggests otherwise we reserve the right to reopen the case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGETING ARM FEMUR GT T2 ALPHA FEMUR ANTEGRADE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18116162
Report Number0009610622-2023-00410
Device Sequence Number7609576
Product Code HSB
UDI-Device Identifier07613327338652
UDI-Public07613327338652
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K203819
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number23533103
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/16/2023
Supplement Date Received by Manufacturer01/04/2024
Initial Report FDA Received Date11/10/2023
Supplement Report FDA Received Date01/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age37 YR
Patient SexFemale
-
-