| Model Number |
71005 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
10/11/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient was implanted with the nalu spinal cord stimulator on (b)(6) 2022 to treat lower back pain.After being implanted the patient is reported to have lost a significant amount of weight and subsequently reports issues with communication between the implantable pulse generator (ipg) and the external therapy discs.A revision procedure took place on (b)(6) 2023.Initial plan was to reposition the ipg within the existing pocket to make it more stable for the patient's current physique.While cutting through the scar tissue to access the ipg, the pocket became enlarged and was no longer usable.The existing pocket was located on the right lower back, a new pocket was created on the left lower back area.While removing the ipg to move it to the new pocket, the device was damaged and became no longer usable.A new ipg and leads were placed.
|
| |
|
Additional Manufacturer Narrative
|
|
There are no allegations of system or component failure, the reported loss of communication appears to be directly related to the patient losing weight and change of physical condition.The ipg was implanted based on the patient's anatomy at the time of implant and was appropriate at that time.
|
| |
|