The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd drive pcb.In addition, our technician confirmed that the electrical pin connector fluid damage, the insertion flexible tube coating damage, the lg connector corroded, the nozzle gluing missing, the air/water socket deformed, the left pulley wire worn out, the right pulley wire worn out, the down pulley wire rupture, and the up pulley wire rupture; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|