| Model Number |
EC-3490TLI |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The time of event is not during procedure.There was no report of patient harm.Video image failure(blackout).
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the image missing.Based on the result, we concluded that it was caused due to the physical damage applied on the ccd module.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|
| |
|