| Catalog Number |
G310SCLR |
| Medical Device Problem Codes |
Overheating of Device (1437); No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
While using a cavitron 300 series g310, they allege that water is not running through and the handpiece is getting hot, no injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Additional Manufacturer Narrative
|
|
Edz dsp/if/ic/scan board shorted screen does not show power setting.Presetting button shows 00 on all of them.Debris buildup in the water filter.Replaced damaged/worn components and recalibrated unit to factory specs.Tech:water flow checked fine 5cc on min and 70cc on max which is good.No sterimate received for evaluation.
|
| |
|