STRYKER GMBH SCALED DRILL Ø2.5MM, AO FITTING; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 703966
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 10/18/2023
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that : "we are contacting you in connection with an incident that occurred in the or during the placement of a plate on a pelvic fracture.Two drill bits from the 2.5mm 300mm ancillary broke during the procedure, remaining in the patient's bone.".
 
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device was returned and matches the alleged failure.The received scaled drill ø2.5mm, ao fitting was visually inspected we can see the cutting flutes are completely deformed, stretch marks are visible on the drill.The surface indicates a brittle fracture caused by strong contact with the fragment after the breakage.The breakage was caused by to application of excessive torsional and axial force.The mark on the cutting flutes indicates an overload fracture.Furthermore, a hardness test according to rockwell on the returned item indicates the material being in accordance with the values in the material specification.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to a user related issue.The failure was caused by an excessive metallic contact which did lead to a mechanical overload.If any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
The customer reported that : "we are contacting you in connection with an incident that occurred in the or during the placement of a plate on a pelvic fracture.Two drill bits from the 2.5mm 300mm ancillary broke during the procedure, remaining in the patient's bone.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCALED DRILL Ø2.5MM, AO FITTING
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18119107
Report Number0008031020-2023-00390
Device Sequence Number15338732
Product Code LXH
UDI-Device Identifier07613252719861
UDI-Public07613252719861
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number703966
Device Lot NumberKU116638
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/18/2023
Supplement Date Received by Manufacturer01/16/2024
Initial Report FDA Received Date11/13/2023
Supplement Report FDA Received Date02/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/18/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-