| Catalog Number |
33680035 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device will not be returned as it remains implanted in the patient.If additional information becomes available, it will be provided on a supplemental report.H3 other text : device remains implanted in patient.
|
| |
|
Event or Problem Description
|
|
It was reported that one of the pegs on the talar dome broke.There is no intervention planned at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed since x-ray images were provided and shows the medial anterior peg on the talar dome is broken.Upon further investigation of the x-ray images by healthcare professionals the following was observed: ¿breakage of the medial anterior peg.There is no relevant sign of loosening or migration visible.¿ more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that one of the pegs on the talar dome broke.There is no intervention planned at this time.
|
| |
|