• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Catalog Number 71335550
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Codes Discomfort (2330); Metal Related Pathology (4530)
Date of Event 11/11/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a plaintiff underwent a right tha surgery on (b)(6) 2009, due to degenerative osteoarthritis, they developed hip pain that threatened mobility and had elevated cobalt levels in serum.A revision surgery was performed on (b)(6) 2021, in which blackish fluid and trunnionosis was found.All acetabular and femoral head components were exchanged for a redapt shell and oxinium s+n implants.No other complications were reported.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the devices were not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, the reported pain, elevated serum cobalt, blackish fluid, and metal debris may be consistent with trunnionosis; however, the source of the reported reactions cannot be confirmed, and it cannot be concluded that the reported clinical reactions were associated with a mal performance of the implant.The patient impact is the reported pain, elevated cobalt, pseudotumor, trunnionosis, and revision.A review of the production orders did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.For all the devices, a review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for total hip systems revealed in warnings and precautions that the patient should be advised to report any pain and unusual incidences.Besides, in the adverse events in primary and revision surgery section it is mentioned that although rare, metal sensitivity reactions and/or allergic reactions to foreign materials have been reported in patients following joint replacement, which have required device removal.A review of the risk management files revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to these products and event.At this time, we have no reason to suspect that the products failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include irregular implant interaction or abnormal motion over time.Based on this investigation, the need for corrective action is not indicated.Should the devices or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 3 HOLE ACET SHELL 50MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18119611
Report Number1020279-2023-02209
Device Sequence Number7540566
Product Code MBL
UDI-Device Identifier03596010598233
UDI-Public03596010598233
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K211176
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/06/2018
Device Catalogue Number71335550
Device Lot Number08GM04990
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2023
Supplement Date Received by Manufacturer01/05/2024
Initial Report FDA Received Date11/13/2023
Supplement Report FDA Received Date01/08/2024
Date Device Manufactured07/08/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EMP 11 STEM PRIMARY SO, LOT#:09HM08394; EMP11 SLV MED CONE1 SPOUT, LOT#:07GAB0019A; HEMI HEAD 38MM, LOT#: 08JW19021; MODULAR SLEEVE -4MM 12/14, LOT#:08GW17829; R3 38MM ID US COCR LNR 50MM, LOT#:09BW22124
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age61 YR
Patient SexMale
-
-