SMITH & NEPHEW, INC. ENDOBUTTON CL ULTRA PAC 1.2; PASSER Back to Search Results
Catalog Number 72202799
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/20/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during acl reconstruction surgery, the endobutton ultra´s guide pin broke off inside patient´s bone.The broken pieces were left inside the patient.It is unknown if there was a back up device available or if there was a surgical delay.No further complications were reported.
 
Additional Manufacturer Narrative
H10: h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states that based on the information provided, the definitive clinical root cause of the reported breakage cannot be determined.The retained broken endobutton ultra´s guide pin is manufactured and intended as externally communicating devices and are not approved for long term internal tissue exposure and long-term implantation data is not available.It was reported the broken guide pin was left inside of the patient¿s bone, therefore, the impact to the patient is the potential for micro-motion/migration, though unlikely as it's contained within the patient¿s bone.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOBUTTON CL ULTRA PAC 1.2
Common Device Name
PASSER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18120992
Report Number1219602-2023-02219
Device Sequence Number6874912
Product Code HWQ
UDI-Device Identifier03596010657626
UDI-Public03596010657626
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K920621
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72202799
Device Lot Number2085198
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2023
Supplement Date Received by Manufacturer12/19/2023
Initial Report FDA Received Date11/13/2023
Supplement Report FDA Received Date12/22/2023
Date Device Manufactured01/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-