• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL PROWLER SELECT LPES; CATHETER, CONTINUOUS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL PROWLER SELECT LPES; CATHETER, CONTINUOUS FLUSH Back to Search Results
Catalog Number 606S155X
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # : (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section h3 - the device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The company is seeking this information through the event investigation.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.This is one of three products involved with the complaint and the associated manufacturer report numbers are 3008114965-2023-00828, 3008114965-2023-00827.
 
Event or Problem Description
As reported by the field, during a coil embolization, an orbit galaxy xtrasoft coil (640cx0406, 30692754) and an orbit galaxy xtrasoft coil (640cx3505, 30706309) became stuck into a prowler select lpes microcatheter (606s155x, 31041133).It was also noted that there is a thread inside the packing of the coil.The procedure was delayed by 30 minutes.The devices were removed easily without additional intervention.A new coil and microcatheter (mc) were used to compete the procedure successfully.There was no report of patient injury.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Complaint conclusion: as reported by the field, during a coil embolization, an orbit galaxy xtrasoft coil (640cx0406, 30692754) and an orbit galaxy xtrasoft coil (640cx3505, 30706309) became stuck into a prowler select lpes microcatheter (606s155x, 31041133).It was also noted that there is a thread inside the packing of the coil.The procedure was delayed by 30 minutes.The devices were removed easily without additional intervention.A new coil and microcatheter (mc) were used to compete the procedure successfully.There was no report of patient injury.A non-sterile prowler select lpes microcatheter was received contained in the decontamination pouch.Upon receiving the device, visual inspection was performed, and no appearance of damages were observed.Microscopic inspection was performed along the body of the prowler, and no damages were found.The device was flushed with a laboratory sample syringe.After that, a 0.013-inch (0.03302 cm) guidewire lab sample was introduced into the received microcatheter, and was able to be advanced through the entire length of the microcatheter without noticeable resistance.The microcatheter was confirmed to be within specifications for the inner diameter (id) and outer diameter (od).A manufacturing record evaluation was performed for the finished device 31041133 number, and no non-conformances related to the malfunction were identified.The guide wire was able to pass through the entire length of the microcatheter without significant resistance, even through the hub.Additionally, no other damages were found on the microcatheter that could have contributed to the issue reported during the procedure.The customer complaint cannot be confirmed with the evidence available.It is possible that other clinical and procedural factors that cannot be replicated during the analysis may have contributed to the reported failure.As part of cerenovus quality process all devices are manufactured, inspected, and released to approved specifications.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no capa activity is required.The instructions for use (ifu) contain the following caution: if strong resistance is met during manipulation, discontinue the procedure and determine the cause of resistance before proceeding.If the cause of resistance cannot be determined, withdraw the catheter and guidewire as a system.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Additional Manufacturer Narrative
Product complaint (b)(4).The product has been returned for evaluation and testing; however, the engineering evaluation has not been completed.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROWLER SELECT LPES
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS INC. (JUAREZ)
circuito interior norte #1820
parque industrial salvarcar
juarez chihuahua 32574
MX   32574
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18121596
Report Number3008114965-2023-00829
Device Sequence Number18018152
Product Code KRA
UDI-Device Identifier10886704028802
UDI-Public10886704028802
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K021591
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2026
Device Catalogue Number606S155X
Device Lot Number31041133
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/26/2023
Supplement Date Received by Manufacturer12/18/2023
12/28/2023
Initial Report FDA Received Date11/13/2023
Supplement Report FDA Received Date12/18/2023
01/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OG CMPLX XTRASOFT COIL 3.5X5; OG CMPLX XTRASOFT COIL 4X6
Patient SexUnknown
-
-