| Model Number |
WA2UR14A |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D1: ureteroscope 7.3/10.4 fr.X 430 mm, 5°, angled ocular, 6.4 fr.Channel.The device was returned to olympus for evaluation, and the customer¿s allegation was confirmed.The evaluation found the following reportable issues: the stopcock closing nut was missing.No other issues were found.The investigation is ongoing and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus that the ureteroscope 7.3/10.4 fr.X 430 mm, 5°, angled ocular, 6.4 fr.Channel had the stopcock come off after the first sterilization.The issue was found during the procedure.There were no reports of patient harm.
|
| |
|
Event or Problem Description
|
|
The intended procedure was completed using the same set of equipment.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Corrected fields: b5 and g2 - health professional is applicable to this event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over one (1) year since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that the missing stopcock closing nut occurred due to excessive use.Olympus will continue to monitor field performance for this device.
|
| |
|