| Model Number |
MI1200 SYNCHRONY PIN |
| Medical Device Problem Codes |
No Device Output (1435); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is suspected.However, to determine an exact root cause a device investigation of the explanted device is necessary.Diagnostic imaging is considered but no date has been scheduled yet.
|
| |
|
Event or Problem Description
|
|
In situ testing showed affected channels and hearing performance is affected.A ct scan is considered.The family and surgeon will be consulted regarding the next steps.
|
| |
|