| Model Number |
VTICMO12.1 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/07/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A4 - unk.A5 - unk.A6 - unk.H6 - work order search: no similar complaint type events were reported for units within the same.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 12.1mm, vticmo12.1, -16.0/2.0/090 (sphere/cylinder/axis), implantable collamer lens was implanted into the patient's left eye (os) on (b)(6) 2023.Lens rotation not associated to a low vault was observed.On (b)(6) 2023 the lens was vertically replaced with a longer length lens and the problem was resolved.Cause of the event is reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: h3 - type of investigation code: 10- device evaluation: the lens was returned in a microcentrifuge vial with moisture.Visual inspection found no visible damage.Dimensional inspection found the lens to be within specifications.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Correction of information: d2b product code: reported as qbc: correct to qcb.D4 primary unique device indenitifer (udi)#: (b)(4), "udi related data quality updates only".Claim# (b)(4).
|
| |
|