It was reported that the patient reported paresthesia to her left lip/chin, status post placement at #18, immediate implant.The inferior alveolar nerve was not seen or palpated during surgery, nor was there any unusual bleeding.Post-op sensation testing revealed a deficit in sensation one week later.The implant was removed.
|
|
This report is being submitted to report additional information.Zimvie received one implant for evaluation.Visual evaluation of the as returned device identified the implant with signs of use, and markings from removal.No damage identified or signs of malfunction that would contribute to the reported event.Measurements match drawing.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are improper techniques used and patient factors.Therefore, based on the available information, a device malfunction did not occur.The reported event was non-verifiable with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|