STRYKER GMBH HANDLE LARGE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 703920
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/11/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
As reported: "turntable at the end of the handle was broken off.".
 
Additional Manufacturer Narrative
Please note correction to section d9/h3, the device was not returned; and h6 (component code).The reported event could not be confirmed, since the device was not returned and no additional information was available.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the lot number was not communicated.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.H3 other text : device disposition is unknown.
 
Event or Problem Description
As reported: "turntable at the end of the handle was broken off.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDLE LARGE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM   D-79111
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18129920
Report Number0008031020-2023-00392
Device Sequence Number12710251
Product Code LXH
UDI-Device Identifier07613327070118
UDI-Public07613327070118
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number703920
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/20/2023
Supplement Date Received by Manufacturer12/14/2023
Initial Report FDA Received Date11/14/2023
Supplement Report FDA Received Date01/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-