MEDTRONIC SINGAPORE OPERATIONS VERSA IPG; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number VEDR01
Medical Device Problem Code Failure to Interrogate (1332)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/27/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that two different programmers were unable to interrogate an implantable pulse generator (ipg).It was also noted that attempting to select the device model prior to finding the patient was unsuccessful.Interrogation was successfully performed using the remote monitoring network.The programmers and ipg remain in use. this event is being reported in response to the advisory describing the potential for device circuit error to occur while the device is processing an atrial-sensed episode. no patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was further reported that the ipg battery status and device history were all in normal range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSA IPG
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SN  486056
Manufacturer (Section G)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SN   486056
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18130364
Report Number3008973940-2023-07383
Device Sequence Number7581151
Product Code NVZ
UDI-Device Identifier00643169709072
UDI-Public00643169709072
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P980035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 11/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/28/2019
Device Model NumberVEDR01
Device Catalogue NumberVEDR01
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2023
Supplement Date Received by Manufacturer11/14/2023
Initial Report FDA Received Date11/14/2023
Supplement Report FDA Received Date11/23/2023
Date Device Manufactured10/08/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
407645 LEAD, 407652 LEAD
Patient SexMale
-
-